Tuesday, March 29, 2016

ICS-CERT Publishes ICS Medical Advisory

This morning the DHS ICS-CERT published their first ‘medical advisory’ for a whole slew (new technical term) of vulnerabilities in older versions of the CareFusion Pyxis SupplyStation. It appears that ICS-CERT is establishing a separate category of advisories for medical devices; the same information just a separate naming-numbering system.

The Advisory


The advisory outlines (saying ‘describes’ would be a gross exaggeration of terminology) over 1400 3rd-party vulnerabilities in older, unsupported versions of the system still running on Windows Server 2003/XP. The vulnerabilities are categorized based upon their CVSS Base score; CVSS 7.0 – 10.0, 715 vulnerabilities; CVSS 4.0 – 6.9, 606 vulnerabilities; and CVSS 0.0 – 3.9, 97 vulnerabilities.

The vulnerabilities were reported by Billy Rios and Mike Ahmadi in collaboration with CareFusion. The affected systems are at the end of their life and CareFusion does not plan on updating the software. The medical advisory does provide a list of mitigating measures that CareFusion recommends owners of the older devices that remain in use.

Commentary


While a new name and number for these medical device advisories will make it easier for medical-device security-researchers to keep up with advisories that specifically apply to their specialties, ICS-CERT is (for now at least) keeping these advisories listed on the same page as the more traditional control system advisories and alerts. I wonder if we are also going to see a new name for other non-traditional controls systems such as building control systems, security control systems and transportation control systems?

Billy and Mike reportedly used an “automated software composition analysis tool” to identify this huge number of vulnerabilities. I’m assuming that what they did was identify the different software components (see the listing in the medical advisory) and then looked up the vulnerabilities listed for each of the components. Then I would assume that CareFusion confirmed that they had issued no patches for any of the listed vulnerabilities. I wonder if they did the same thing for the various 3rd-party libraries used by the various components? I do hope that Billy and Mike are planning on doing a presentation on this investigation at one of the conferences this year.

This is, HOPEFULLY, an outlier example of the problem of bundling software systems and then attempting (or not attempting as is apparent in this case) with keeping up with all of the patches and updates for the various programs. This may be an extreme example, but I would be that it is not the only product with this problem, either in the medical device category or any other ICS category for that matter.


As a side note, don’t be too quick to chide the Pyxis owners for still using medical devices base upon the outdated Server 2003/XP products. CareFusion has apparently worked with Microsoft to continue supporting their systems past the normal XP end-of-life. Apparently smaller hospitals and other medical facilities with limited equipment budgets use an equipment utilization model that is not uncommon in more traditional automation environments; if it ain’t broke, don’t mess with it.

Wednesday, October 15, 2014

ICS-CERT Updates Bash and Siemens OpenSSL - Issues Pyxis Advisory

Today the DHS ICS-CERT published four documents concerning vulnerabilities in control systems. One is actually a correction to yesterday’s Siemens OpenSSL update. Then they updated the BASH advisory and issued a supplement to that advisory. Finally they published a new advisory for a medical supply system with multiple vulnerabilities.

Siemens OpenSSL Update

Yesterday ICS-CERT published an update for the Siemens OpenSSL advisory that reported that all affected systems had updates available. Today they are reporting that what Siemens actually said was that a new update was available for ROX 2-based devices. And it reports that Siemens is still working on an update for ROX 1.

Oops, no, they got it wrong again. According to the Siemens ProductCERT advisory, the newest update only affects ROX V2.6.0 with Crossbow V4.2.3.  There are three products listed by Siemens that do not currently have updates available:

• APE 1 with eLAN installed: All versions <= 1.0.1;
• ROX 1: All versions (only affected if Crossbow is installed);
• ROX 2 with eLAN installed: All versions < V2.6.0

Sorry. I did not read the Siemens ProductCert advisory on this yesterday; I trusted ICS-CERT to get it right. Well, maybe they’ll get it right tomorrow. And that will be version ‘F’ when we should still be on version ‘C’. Oh well….

BASH Advisory Update and Supplement

Back in September ICS-CERT published a brief advisory on the Bash command injection vulnerability. I was kind of busy at the time and didn’t write about their advisory because much more complete information was readily available elsewhere. Well, today the published an update to that advisory that was not much better. The only change was the addition of the following paragraph:

“ICS-CERT sent out a query to vendors we have collaborated with in the past. Many have responded back with information about which products are affected by this bash vulnerability. ICS-CERT created a supplement to this advisory that contains this information. It can be found at the following web location: https://ics-cert.us-cert.gov/advisories/Supplement-ICSA-14-269-01. This supplement will be updated with additional information as it becomes available, without updating this advisory.”

So now we have a new ICS-CERT document that will be periodically updated so they don’t have to update the advisory so often??? Okay, what ever.

Okay, the supplement provides some useful information. First it provides a list of companies that have responded to ICS-CERT inquiries about potentially vulnerable systems. Then it provides a list of vulnerable systems (by vendor) with links to further information. It is not clear that systems from vendors on the first list that are not listed on the second list are actually not vulnerable. I think that that may be a dangerous assumption to make. In any case, selected products from the following vendors are reportedly at least potentially vulnerable:

• ABB;
• Cisco;
• Digi;
• eWON;
• Meinberg;
• Moxa;
• Red Lion (pardon me; use bash shell but “are not considered to be vulnerable or exploitable”; and
• Siemens (okay, this lets them off the hook for not mentioning the Siemens ProductCERT advisory yesterday).

Pyxis Advisory

NOTE: This is for a medical supply control system, not really an industrial control system, but hey the FDA won’t touch this and ICS-CERT doesn’t have anything else going on right now….

This advisory is for multiple authentication vulnerabilities in the CareFusion Pyxis SupplyStation reported by Billy Rios. CareFusion has produced a new version of the software that mitigates three of the four vulnerabilities. No mention if Billy was given the chance to verify the efficacy of the fix.

ICS-CERT reports that the vulnerabilities are:

• Hard-coded password, CVE-2014-5422;
• Hard-coded credentials, CVE-2014-5421 and CVE-2014-5420; and
• Insecure temporary files, CVE-2014-5423

ICS-CERT reports that a relatively low skilled attacker could remotely exploit these vulnerabilities to manipulate the locking controls on the automated medical supply cabinets. I don’t expect that these are used to dispense narcotics or else the DEA would be involved.


CareFusion reportedly will not be offering a fix to one of the hard-coded credential vulnerabilities because it would only allow access to some application files, but not the physical access controls. That would seem to be a reasonable risk assessment decision if it was made by the system owner, not the manufacturer. Good news, it will be fixed in future versions.
 